Biolis 24i Premium Support

Controls are required daily.
Does the Biolis need calibrating?
Yes, if daily control gives readings which are out of range.
How do I store the reagents(chemistries) when I am not using the analyser?
Reagents can be left on the Biolis 24i thanks to the on-board refrigeration unit.
What maintenance is required?
Minimal – simple tasks such as emptying the waste, checking levels of water and washes and cleaning of the filters will need to be done on a regular basis to maintain optimal analyser performance
What service/support is provided?
We have a team of engineers that provide phone support plus we do routine on-site servicing at regular intervals all provided under your rental agreement at no extra charge.